Model:FT-SZZL
Brand:tianqiong
1.Total Phosphorus Analyzer Principle
Total Phosphorus Analyzer is used to monitor total phosphorus levels, offering a wide measurement range suitable for a variety of wastewater and drinking water applications.Under high temperature and high pressure conditions, the sample is digested with potassium persulfate, and all the phosphorus-containing compounds in the sample are converted into orthophosphate (this step is required to measure orthophosphate).In a 5%-8% nitric acid solution, orthophosphate forms a soluble phosphovana-molybdenum yellow complex with ammonium metavanadate and ammonium molybdate, and its absorbance is measured at a specific wavelength, which is linearly related to the orthophosphate concentration of the sample.In this relation, the absorbance can be converted into the total phosphorus (TP) or orthophosphate (PO4) concentration value of the sample.
2.Total Phosphorus Analyzer product features
The new optoelectronic quantitative technology is not affected by interference such as large chromaticity, many suspended objects, and many bubbles.It has high reliability and accuracy and strong anti-interference ability.
High detection accuracy, low measurement limit, and low long-term drift of the instrument.
It has one-click self-test function and self-diagnosis function.
It adopts a single-channel high-integrated valve set, making maintenance and cleaning simple and fast.
All-weather networking function, monitoring the operating status of the instrument anytime and anywhere.
It has alarm function, quality control function and counter control function.
It has automatic cleaning function and automatic calibration function.
It has full-point measurement, interval measurement and external control measurement.
It works automatically after power failure recovery.
Automatic prompts and automatic reset functions for faults and reagents (samples).
It has fault logging function.
Can save historical data for more than 3 years.
Satisfies the separation of cleaning liquid and waste liquid
3.Technical parameters of Total Phosphorus Analyzer
Measurement method: ammonium molybdate spectrophotometry (HJ670-2013)
Measuring range: 0-2 mg/L; 0-10 mg/L; 0-50 mg/L; range can be customized
Display error: ≤±3% detection limit: 0.010mg/L
Repeatability: ≤±3% Waste liquid volume: 4.5mL/time
Memory effect: ≤±1%Real water sample comparison: ±8%
Constant temperature time: 900 seconds Voltage stability: ≤±5%
Measurement cycle: 39 minutes Calibration cycle: any specified time
Maintenance cycle: ≥720h/time Maintenance workload: <2 hours/month
Reagent consumption: 3 months/500ml display output: configured with a 10.1-inch color LCD touch screen
Color rendering temperature: The recommended temperature is 120℃, which can be set according to the actual water sample condition
Data export: The measured value can be imported into the USB drive and saved in the USB drive.
Signal output: RS485/RS232/USB interface/standard 2-channel 4-20mA output/standard two-channel switching input and output
Environmental requirements: Indoors with controllable temperature, recommended temperature (5-28℃), humidity ≦90% (no condensation)
Power supply and power: (220V±22) V/AC, (50±0.5Hz), 5A, 150W
Instrument size: upper cabinet 600*450*300mm; upper cabinet 700*450*300mm
During ship navigation, when encountering waters with unknown water depth conditions, ship drivers face potential risks such as reef striking and stranding, which seriously threaten navigation safety. In such cases, a Portable Ultrasonic Water Depth Meter can be used.The Portable Ultrasonic Water De...
In the production and use of photovoltaic modules, grid breakage is a relatively common and thorny problem. Grid breakage refers to the existence of breakpoints in the silver paste printing of the main grid of the solar cell, which causes the fine grid lines on the solar cell to be open-circuited wi...
The "data brain" in farm fields is quietly changing the agricultural production model that has relied on "weather-dependent farming" for thousands of years.In traditional agricultural production, farmers mostly rely on experience to judge agricultural activities. However, as clim...
Against the backdrop of accelerating informatization in water conservancy, the demand for monitoring in the field of water conservancy management is rising increasingly. To address issues such as frequent flood disasters and the need to improve the efficiency of water resource management, a Water Le...