1.Water Hardness Tester product introduction
Water Hardness Tester is designed to quickly and accurately measure water hardness (total water hardness and chloride ion) to ensure hardness levels remain within acceptable ranges.GL02 Online Water Hardness Tester is an instrument that can monitor the total hardness and chloride ions online.It integrates water hardness sensor, chloride ion sensor and measurement module communication storage, and can quickly and accurately record key parameters in water.At the same time, the instrument supports extended water quality multi-parameter sensors, including but not limited to pH, dissolved oxygen (DO), conductivity, turbidity, temperature, ammonia nitrogen, etc., which can be combined and configured according to different needs and applications, record and store historical monitoring data and alarm history records, and support the export of historical data.xlsx.The RS485 interface supports MODBUS-RTU communication protocol to facilitate users to communicate freely with PLC, DCS, configuration software, DTU and other devices to connect and transmit data.
2.Water Hardness Tester application field
Online Water Hardness Tester is widely used in the monitoring and control of various water bodies, including but not limited to the following areas:
1.Factory boiler water quality monitoring: used to prevent scale formation, prevent equipment corrosion, and ensure equipment safety.
2.Groundwater monitoring: used to monitor the pH, conductivity, temperature and other parameters of groundwater in order to promptly discover and solve water quality problems.
3.River and lake monitoring: used to monitor the water quality of rivers and lakes, such as dissolved oxygen, turbidity, ammonia nitrogen and other parameters, so as to take pollution control measures in a timely manner.
4.Ocean monitoring: used to monitor the ocean's water quality, such as salinity, dissolved oxygen, temperature and other parameters, so as to promptly detect and control marine pollution.
5.Sewage treatment: used to monitor the water quality parameters of sewage, such as pH, COD, ammonia nitrogen, etc., in order to control and regulate the sewage treatment process.
6.Industrial production: used to monitor the water quality conditions in the industrial production process, such as pH, conductivity, dissolved oxygen and other parameters, so as to adjust the process process in a timely manner and ensure product quality.
7.Water plant: used to monitor the pH value, dissolved oxygen, turbidity and other parameters of tap water to ensure the safety and hygiene of tap water.
8.Scientific research: used for water quality monitoring in the field of scientific research, such as lake eutrophication, climate change and other research.
3.Water Hardness Tester technical features
1.High reliability: suitable for long-term working in the wild environment, stable measurement and strong anti-interference ability.
2.Flexible and portable: Each probe can be freely combined, replaced independently, and used for plug and play.
3.Scalability: Freely combine multiple sensors.
4.Multiple applications: rapid on-site measurement, emergency monitoring, or long-term online monitoring of groundwater, river water, lake water sources, and urban pipeline water.
5.Tough shell: ABS+PC material, corrosion-resistant, can work normally for a long time.
6.Compact structure: can be installed in smaller sizes.
7.Communication connection: RS485 extended interface, master/slave interface isolation can communicate independently.
4.Water Hardness Tester technical parameters
Display output | 4.3-inch touch screen with LED strong backlight, can be operated under direct sunlight |
power supply | DC power supply: DC12V |
Power consumption | The instrument power consumption is about 12V/1W |
Sound output | buzzer |
Communication protocol | Standard RS485 Modbus-RTU protocol and device master/slave transmission channel support |
Main materials | ABS+PC material |
Storage temperature | -20 to 70℃ |
Operating temperature | -10 to 50℃ |
Protection level | IP65 |
size | 175mm*140mm*49mm(Length×Width×Height) |
weight | About 0.5KG |
5.Water Hardness Tester sensor configuration (optional)
Serial number | name | Measurement range | principle | Measurement accuracy | Configuration | Remark |
1 | temperature | 0~50℃ | High-precision digital sensor | ±0.3℃ | ||
2 | pH | 0~14 (ph) | Electrochemistry (salt bridge) | ± 0.1PH | ||
3 | ORP | -1500mv~1500mv | Electrochemistry (salt bridge) | ± 6mv | ||
4 | Conductivity | 0~5000uS/cm | Contact electrode method | ± 1.5% | ||
5 | Turbidity | 0~40NTU (low turbidity) | Scattering light method | ± 1% | Optional | |
0~1000NTU (medium turbidity) | Scattering light method | ± 1% | ||||
0~3000NTU (high turbidity) | Scattering light method | ± 1% | ||||
6 | Dissolved oxygen | 0~20mg/L | Fluorescence life method | ± 2% | ||
7 | Ammonia nitrogen | 0~100mg/l | Ion Selection Electrode Method | ±1mg | ||
8 | Suspension | 0~2000mg/L | Scattering light method | ±5% (depending on sludge homogeneity) | ||
9 | Residual chlorine | 0~5.00 mg/L | Ion Selection Electrode Method | ±5% of reading | ||
10 | Chloride ions | 0-3500.0mg/L | Ion Selection Electrode Method | ±5% | ✔ | |
11 | Total hardness | 0~1000.0mg/L | Ion Selection Electrode Method | ±10% of reading | ✔ | |
12 | cod | 0~500mg/ L | UV254 absorption method | ±5% | COD and chlorophyll have automatic cleaners, which can prevent biological adhesion and avoid light window contamination, so as to ensure that long-term monitoring is still stable; automatic cleaning time and cleaning times can be set, and the power consumption is 0.7W | |
13 | Chlorophyll | 0~400ug/L | Fluorescence method | R2>0.999 |
The Ultrasonic Wind Speed Direction Sensor achieves measurement based on the characteristics of ultrasonic waves propagating in the air. It accurately calculates the wind speed and direction by measuring the time difference of ultrasonic waves propagating in different directions. This non-mechanical...
The microclimate of farmland refers to the small-scale climatic environment formed by the interaction among the surface air layer, soil layer and crop population in the farmland. It is reflected by the values of agricultural meteorological elements such as radiation, air temperature and humidity, wi...
In the context of the increasingly accelerating urbanization process today, air quality has become one of the key elements for measuring the quality of life. However, due to the high construction and operation costs and fixed layout, traditional large-scale environmental monitoring stations find it...
In the modern industrial field, the chemical industry has an extremely high demand for meteorological monitoring. The chemical production process is complex and involves many flammable, explosive, and corrosive chemical substances. For example, in a chemical industrial park, under different meteorol...